Vigor LC Nitrile Exam Gloves Long Cuff

View Product

> Fingertip textured
> Powder Free
> Not made with natural rubber latex
> Chemotherapy drugs tested
> UL certified to meet NFPA 1999
> Ambidextrous
> Long cuff
> Orange colour

 


Longer in cuff length, GloveOn Vigor LC is designed for robustness and hardiness to get the tough jobs done. The vibrant choice of orange highlights its uniqueness and suitability for industrial and hospitality markets.

 

Reorder CodeSize
NTR27SSSmall
NTR27MMMedium
NTR27LLLarge
NTR27XLX-Large

 

Glove PropertiesMeasurement (mm)
Length≥280
Thickness at Palm (centre of Palm)0.13 ± 0.02
Thickness at Finger (13mm ± 3mm from tip)0.17 ± 0.02
Physical PropertiesBefore AgeingAfter Ageing
Tensile Strength (MPa)≥18≥16
Elongation (%)≥500≥500
Elongation after Isopropanol immersion (%)≥500N/A
Inspection Levels and AQLInspection LevelAQL
WatertightnessG11.5
Physical DimensionsS24.0
Tensile StrengthS24.0
Visual Inspection (Major)S42.5
Visual Inspection (Minor)S44.0
Particulate ResidueN=5≤2mg/glove
Packaging 
Configuration100 gloves per box
10 boxes per carton
Regulatory Compliance 
RegulationsTGA - ARTG 164563
FDA 510(k)
MDD 93/42/EEC
REACH
ROHS Directive 2002/95/EC
International Standards 
StandardsASTM D6319
ASTM D6978
EN 455 part 1, 2, 3, 4
NFPA 1999
Manufacturing Accreditations 
AccreditationsISO 9001:2015
ISO 13485:2003
EN ISO 13485:2012
Chemotherapy Drugs and Concentration
(Tested for Resistance to Permeation by Chemotherapy Drugs as per ASTM D6978-05, Test Report PN 104145A)
Minimum Breakthrough Detection Time (minutes)
Carmustine (BCNU), 3.3mg/ml (3,300ppm)45.5
Cisplatin, 1.0mg/ml (1,000ppm)>240
Cyclophosphamide (Cytoxan), 20.0mg/ml (20,000ppm)>240
Dacarbazine (DTIC), 10.0mg/ml (10,000ppm)>240
Doxorubicin Hydrochloride, 2.0mg/ml (2,000ppm)>240
Etoposide (Toposar), 20.0mg/ml (20,000ppm)>240
Fluorouracil, 50.0mg/ml (50,000ppm)>240
Methotrexate, 25.0mg/ml (25,000ppm)>240
Mitomycin C, 0.5mg/ml (500ppm)>240
Paclitaxel (Taxol), 6.0mg/ml (6,000ppm)>240
Thiotepa, 10.0mg/ml (10,000ppm)30.9
Vincristine Sulfate, 1.0mg/ml (1,000ppm)>240
WARNING: Carmustine and Thiotepa, at the tested concentration, degraded a Vigor LC nitrile glove at 45.5 minutes and 30.9 minutes respectively. The safe use of gloves in chemotherapy treatment is solely the decision of clinicians authorised to make such a decision.